Dynamic Simulation of Moving Element for Dynamic Calibration Device of Eddy Current Sensor Based on ANSYS

Xiao HE, Xiang XU, Sen YANG

Abstract


In order to ensure that the moving element of dynamic calibration of the eddy current sensor does not resonate and the unevenness of vibration does not affect the accuracy of calibration, modal analysis and harmonic response analysis was performed within the frequency range of the dynamic calibration device of the eddy current sensor. The results show that the first order modal frequency of the moving element is 3406 Hz, which is 5 times greater than the maximum calibration frequency. The vibration unevenness of the induction plate at maximum working frequency is 0.73%, which does not enough to significantly influence on the calibration accuracy. Therefore, the design of the moving element of dynamic calibration device for the eddy current sensor meets the requirements of operation.

Keywords


Dynamic calibration, Moving element, Induction plate, Modal analysis, Harmonic response analysis